咨询服务热线:
60593951/13701191584
Product Service Products
Diesel engine emission particle weighing box
Time:2021-10-16 Source:
Share

model: PW +20/380-1065


Adopting standards:


China: GB18352.6-2016,GB 17691-2018
U.S.A: EPA 40CFR86,EPA US 2010(CFR 1065) 
Europe: 2003/76/EC, 715/2007/EC, 692/2008/EC
The particle weighing box of EST complies with all 7 standards of the US EPA US 2010 (CFR 1065). Control software and controller developed by EST company
Shock absorber device designed and manufactured by EST company
Temperature control and humidification system designed and manufactured by EST company
Environmentally friendly refrigerant
EST air purification system that meets the latest standards
Manufacturing standard: Chinese National Standard
The seven requirements of EPA Regulation 1065 for particle weighing bins in the United States: (see § 1065.190)

Serial Number

Parameter

Parameter values

1

Dust prevention

The weighing environment meets the cleanroom standards: ISO14644-1 Class six

2

windbreak

Wind speed in weighing area: ≤ 0.05 m/s

3

constant temperature

Weighing environment temperature: (22 ± 1) ℃

4

dew point

Dew point of weighing environment: 9.5 ℃± 1 ℃

5

pressure monitoring

Weighing environment pressure detection: ± 100 Pa

6

shockproof

Isolation platform isolates external noise and vibration

7

Anti static

Weighing environment electrostatic field: ≤ ±2V


 

Basic configuration:

 

Particle weighing box body
Temperature and humidity dew point control system
Breeze regulation control system
Shock absorber device
Clean control system
Control and Monitoring System
Static elimination system

 

Technical parameters:



Parameter

Experimental volume

380 liters

External dimensions

1720 x 820 x 2000 mm (W*D*H)

Internal dimensions

890 x 630 x 680 mm (W*D*H)

Outer panel of the box

Galvanized steel plate, spray coating treatment

Equipment entrance and exit doors

Twisted chain on top

Observation window

660 x 660 mm (W*H)

Side cleanliness transmission window

250 x 250 mm (W*H)

Operating hand hole

ϕ190mm, Sealed operating glove assembly

Inner surface

SUS304 stainless steel

Floor load-bearing capacity

100KG

Filter paper storage rack

SUS304, Maximum load 10KG

lighting

Inside the experimental chamber


 

Temperature and humidity control system inside the experimental chamber:
 

Temperature control:

 

Name

Parameter

temperature range

+20 ℃ to +30 ℃

temperature fluctuation

±0.5 ℃

Temperature display accuracy

±0.1℃

refrigeration

Directly cooled, adjustable, refrigerant R134a

Heating

Stainless steel heater with overheat protection

 

湿度控制:                     

 

Name

Parameter

temperature range

+20 ℃ to +30 ℃

Humidity range

Relative humidity ranging from 40% to 50%

Humidity fluctuation

Relative humidity ± 3%

Humidity display accuracy

± 1% (steady state)

     

 

Dew point range control: 9.5 ℃± 1 ℃

The humidifier water can be automatically replenished through a water storage tank. Automatic circulation of condensate water.

The air circulation circuit has been specially designed. Measurement area wind speed ≤ 0.05m/s

The above data was obtained under fixed environmental conditions: temperature+25 ℃, power supply voltage 220V, 50Hz, no thermal load

Environmental conditions for use: Temperature range:+15 ℃ to+35 ℃, no condensation.

 

Temperature and humidity sensors:     

 

VAISALA HMT330 series temperature and humidity transmitter

Name

Parameter

Temperature measurement accuracy

±0.2℃

Temperature measurement resolution

0.1℃



Relative humidity accuracy

At+15+ At 25 ° C:

              ±1 %RH (0 ... 90 %)

              ±1.7 %RH (90 ... 100 %)

At+25+ At 40 ° C:

               ±(1.0 + 0.008xreading) %RH


 

Shock absorber device:

 
The function of shock absorbers is to prevent external vibrations from being transmitted to the microbalance and causing an unstable state.

EST Micro Balance Shock Absorption System

 

Name

Parameter

Dual stage damping and shock absorption platform

Natural stone + metal platform

Metal platform size

730 * 560mm (W*L)

Platform load

20kg

natural frequency

≤10Hz

Mechanical shock absorber

Special design

weight

>20kg

Stone platform size

200 * 350mm (W*L)

Platform load

5kg

Mechanical shock absorber

Special design

 

洁净度保持系统:

 

If the laboratory environment conditions cannot reach the cleanliness level, the test chamber must be equipped with a suitable high cleanliness filter to filter the air inside the chamber. In addition, a transition window for air shower transmission with high cleanliness filtration should be installed to prevent dust from entering and exiting the cabin with the filter box.

 

Name

Parameter

Cleaning grade

Meets and exceeds the US federal EPA standards

Transition window for air shower transmission

Installed on the right side bulkhead

Transfer window size

250 * 250mm

Two electric doors for one in and one out

High cleanliness air shower filter

Control the internal and external switches separately

High cleanliness filter

Can be replaced with a high-efficiency filter without partitions

Filter size 630 * 490 * 46mm (width * depth * height)

Booster fan

Flow control valve

Diversion net

 

Voltage regulator box
Built in booster fan
Built in pre filter (medium cleanliness filter)

 

Pressure monitoring device

 

Name

Parameter

Differential pressure gauge
(Used to monitor the pressure difference of the filter inside the test chamber)

Measurement range: 0-60Pa
Resolution: 1Pa
Accuracy: 2.0% of pt or 1% of max


High precision absolute pressure sensor
(Used to measure the ambient air pressure inside the environmental chamber)

Capacitive quartz crystal material
Measurement range (range): 90-110kPa
Accuracy: ± 0.2%
Accuracy ± 75Pa, repeatability ± 25Pa, noise ± 5Pa

Static elimination system:

 

The board inside the test box is made of excellent conductive metal and well grounded
Filter paper storage rack with excellent conductive metal and good grounding
DC micro ion fan working power supply: 220V/50Hz
                         Operating temperature: 0-50 ℃
                         Dimensions: 50mm (D) × 78mm (W) × 113mm (H)
                         Static voltage: ± 2V
                         Placement in the testing chamber
                         Stainless steel normally closed tweezers
                         Grounding environment: power supply grounding and shell grounding for the warehouse

 

Control and monitoring system:

 

Monitor              Color LCD display

 

Monitor              the set temperature value and the actual value
                        Humidity setting value and actual value
                        Over temperature protection
                        Warehouse pressure
                        Fault code display
                        running state 

Control              the operation terminal at the front of the experimental chamber
                        Touch screen setting values
                        The pressure difference instrument inside and outside the box is located on the front of the box

 

External device communication interface                network interface

 

The manufacturer's program fully complies with the safety and personal health protection directives of China and Europe

Chinese standards:
-GB 150-1998 Steel Pressure Vessel Standard
-GB 4343-1995 EMC standard
-GB5226.1-2002 Mechanical Safety Mechanical Electrical Equipment Standard